Medical and industrial laboratories now have two economical options for producing laboratory grade and Type I ultrapure water for analytical, life sciences and general lab purposes. Both point of use lab water systems produce Type I ultrapure water that meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
Lab Water Purification System (Model 0021-0024)
The Lab Water Purification System produces ultrapure water from reverse osmosis (RO), distillation or deionized water. The system includes a purification cartridge, 0.2 micron PES capsule filter and a 185/254nm dual wavelength UV lamp.
Features
- Automatic system recirculation every hour to prevent bacteria from growing inside the system
Large capacity ultra-purification cartridges contain uniquely formulated ion resin and fine activated carbon to ensure long useful life and a low running cost
A 0.2 micron final filter removes any remaining contaminants
A dual UV lamp eliminates bacteria and reduces organics to trace level
Volumetric dispensing and calibration for easy dispensing of ultrapure water
Real time resistivity measurement by a high precision resistivity meter (conductivity cell constant 0.01cm-1)
Alarms when consumables reach the end of life
Small compact footprint occupies minimal space and is easy to move around with an ergonomically designed handle
Direct Lab Water Purification System (Model 0021-0023)
The Direct Lab Water Purification System is an integrated system which can produce ultrapure and reverse osmosis (RO) water directly from tap water. System includes a 3-stage prefiltration kit, RO membrane, 30L PE water tank with level sensor, tank vent filter with CO2 remover, high purity water filter cartridge, ultrapure water filter cartridge, 0.2 micron capsule filter and a 185/254nm dual wavelength UV lamp.
Features
- Prefiltration pack removes large particles, calcium and organic compounds, protecting RO membrane for extended life
Fully automated control system has pre-rinse, RO membrane fouling prevention flush, RO membrane and cartridge life detection
Automatic RO membrane rinse for new membranes
Two sets of dual-column purification cartridges work sequentially to ensure thorough removal of trace ions and organics in water
Programmable water dispensing for ease of water collection
A 0.2 micron final filter removes any remaining contaminants
The PE tank with conical bottom ensures no dead space to prevent bacteria growth
A dual wavelength UV lamp kills bacteria and reduces organics to trace level
System removes over 99% of large molecules and particles and 95% of ions in water
- Resistivity is measured by a high-precision resistivity meter (conductivity cell constant 0.01cm-1)
LAB SPECIFICATIONS
- Operating Voltage: 110V
Operating Temperature: 5-45°C
Feed Water Pressure*: 0-15 PSI (0-1 Bar)
Production Rate at 25°C: 1.0-1.5 L/min
Resistivity at 25°C: 18.2 MΩ.cm
TOC Level**: <5 ppb
Particles (>0.2 micron): <1/mL
Microorganisms: <1cfu/mL
Pyrogens (Endotoxin): <0.001 EU/mL
Dimensions: 8″W x 13 3/4″D x 17 3/4″H
DIRECT LAB SPECIFICATIONS
- Operating Voltage: 110V
Feed Water: Municipal Water TDS <1,000 ppm
Operating Temperature: 5-45°C
Feed Water Pressure*: 0-90 PSI (0-6 Bar)
Production Rate at 25°C: Type I water 1.2 L/min, RO water 10 L/h
Resistivity of Ultrapure Waterat 25°C: 18.2 MΩ.cm
Ultrapure Water TOC Level**: <5 ppb
Particles in Ultrapure Water(>0.2 micron): <1/mL
Microorganisms: <1cfu/mL
Pyrogens (Endotoxin): <0.001 EU/mL
System Dimensions: 13″W x 19″D x 20″H
Tank Dimensions: 16″ Dia x 26″H
RO Rejection Rate: 95%
Water Tank Capacity: 30 L